45
Blood pressure
Input/output impedance
350 Ω
Exciter input limit ± 10 V
Exciter input frequency
range
DC to 4000 Hz
Transducer sensitivity 5 µV/V/mmHg or 40 µV/V/mmHg
Level accuracy ± 1 %, ± 1 mmHg
Static levels BP1 -10 mmHg, 0 mmHg, 50 mmHg, 100 mmHg, 150 mmHg,
200 mmHg, and 250 mmHg
Static levels BP2 -10 mmHg, 0 mmHg, 80 mmHg, 160 mmHg, 240 mmHg,
320 mmHg, and 400 mmHg
Channel selections: Arterial 120/80, channel 1 and 2
Radial artery 120/80, channel 1 and 2
Left ventricle 120/00, channel 1 and 2
Right ventricle 25/00, channel 1 and 2
Central venous 15/10, channel 2
Pulmonary artery 25/10, channel 2
Pulmonary wedge 10/2, channel 2
Left atrium 14/4; automatic Swan/Ganz (every 20 sec)
Manual Swan/Ganz (changes when entry is selected), channel 2
Synchronized with all normal sinus rates
Physiologically track all arrhythmia selection
Cardiac output (must have optional cardiac output adapter box 2651740)
Catheter type Baxter edwards, 10 cc
Blood temperature 37 °C (98.6 °F)
CO for 2 °C (35.6 °F): 3, 5, 7 l/min
CO for 20 °C (68 °F): 3, 5, 7 l/min
Cal pulse
Of 1 °C for 1 sec; of Delta 402 Ω for 4 sec
Computational constant For 2 °C (35.6 °F) is 0.561; for 20 °C (68 °F) is 0.608
Left to right shunt 2 °C and 20 °C (35.6 °F and 68 °F)
Faulty injectate 2 °C and 20 °C (35.6 °F and 68 °F)
Accuracy ± 5 %
Calibrated or uncalibrated cardiac output waves for 4 different CO
values
Respiration
Baseline impedance
500 Ω, 1000 Ω, 1500 Ω, and 2000 Ω, leads I, II, and III
Lead selections LL or LA
Impedance variations
3 Ω, 1 Ω, 0.5 Ω, and 0.2 Ω
Accuracy ± 5 %
Rates 15 BPM, 20 BPM, 30 BPM, 40 BPM, 60 BPM, 80 BPM,
100 BPM, 120 BPM, and 0 BPM for apnea
Accuracy ± 2 %
Apnea 12 seconds, 22 seconds, 32 seconds, and continuous
Temperature
Compatibility YSI 400/700 series
Temperature 30 °C, 35 °C, 37 °C, 40 °C, and 42 °C
(86 °F, 95 °F, 98.6 °F, 104 °F, and 107.6 °F)
Temperature
simulation accuracy
± 0.25 °C
PS420
Patient Simulator
Specications
Patient Simulators