Dynatech MP-7000 Manuale di Servizio Pagina 39

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 38
Functional Verification Tests
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 35
Temperature Test
Tools Needed IBP Temp Simulator, or equivalent
Test Setup None
Do this... Observe This... Pass/Fail
1. Connect the temperature simulator to the
Propaq MD T1 channel.
2. On the IBP Temp Simulator, Select 98.6
Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.
The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.
oo
3. Repeat steps 1 and 2 of this procedure for the
T2 channel.
Vedere la pagina 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 91 92

Commenti su questo manuale

Nessun commento